Abstract
Lab-on-a-chip (LOC) devices for electrochemical analysis of DNA hybridization events offer a technology for real-time and label-free assessment of biomarkers at the point-of-care. Here, we present a microfluidic LOC, with 3×3 arrayed electrochemical sensors for the analysis of DNA hybridization events. A new dual layer microfluidic valved manipulation system is integrated providing controlled and automated capabilities for high throughput analysis. This feature improves the repeatability, accuracy, and overall sensing performance (. Fig. 1). The electrochemical activity of the fabricated microfluidic device is validated and demonstrated repeatable and reversible Nernstian characteristics. System design required detailed analysis of energy storage and dissipation as our sensing modeling involves diffusion-related electrochemical impedance spectroscopy. The effect of DNA hybridization on the calculated charge transfer resistance and the diffusional resistance components is evaluated. We demonstrate a specific device with an average cross-reactivity value of 27.5%. The device yields semilogarithmic dose response and enables a theoretical detection limit of 1. nM of complementary ssDNA target. This limit is lower than our previously reported non-valved device by 74% due to on-chip valve integration providing controlled and accurate assay capabilities.
Original language | American English |
---|---|
Pages (from-to) | 579-585 |
Number of pages | 7 |
Journal | Biosensors and Bioelectronics |
Volume | 64 |
DOIs | |
State | Published - 5 Feb 2015 |
Externally published | Yes |
Keywords
- DNA hybridization sensing
- Electrochemical impedance spectroscopy
- Label-free detection
- Microfluidics
- Restricted diffusion
- Valve
All Science Journal Classification (ASJC) codes
- Biotechnology
- Biophysics
- Biomedical Engineering
- Electrochemistry